Procedure for Dialysis Machine
Disinfection
Daily
Machine Disinfection:
- Substitute undiluted Alcavis 100
for bleach.
- Pull 200 - 250 cc of Alcavis 100
into the dialysis machine through the dialysate line
or bicarbonate line. Do not use acid connector/line.
- The machine will dilute Alcavis
100 1:34 with water. This results in an available chlorine
of 333 ± 10 ppm.
- Confirm presence of Alcavis 100
with a test strip from the drain line.
- After drawing 200 - 250 cc of Alcavis
100, put the machine in "RINSE" procedure
for 10 - 15 minutes. (Up to a 5 minute dwell period can
be performed, but is not required).
- Confirm absence of Alcavis 100
with a test strip from the drain line.
- If chlorine is detected (= 0.5 ppm),
repeat the "RINSE" procedure.
- Never let Alcavis 100 dwell
in the machine for more than 20 minutes.
up
Procedure for the Disinfection
of the "WHO"
Waste Handling Option (WHO) Disinfection:
- Using a syringe and a WHO priming
connector, inject 15 cc of Alcavis 100 into the WHO.
- Remove the syringe and priming connector
and replace the WHO rinse arm. The WHO rinse arm must be
replaced, otherwise proper rinsing of the WHO will not occur.
- Allow the machine to rinse for a minimum
of 5 minutes prior to setting up for the next patient.