RINSING/CLEANING |
ARM-01 rinses
both the blood and dialysate compartment
with AAMI quality water along with a
cleaning solution such as peracetic
acid / hydrogen peroxide. |
TOTAL
CELL VOLUME TEST |
The dialyzer
is tested to determine the total cell
volume Pass/Fail criteria and is based
upon the percent of the original priming
volume according to the clinic's protocol
(>80% minimum). |
PRESSURE
LEAK TEST |
The dialyzer
blood side compartment is pressurized
with the dialysate compartment vented
to drain. The rate of leak (mmHg / minute)
is calculated and the appropriate pass
/fail criteria are applied. |
DISINFECTANT
FILL |
The dialyzer
under test will be filled with the chemical
germicide to destroy and retard bacterial
growth. |
DIALYZER
LABELING |
Each reprocessed
dialyzer will have a unique label that
allows accurate device and patient identification. |
CHEMICAL
AGENTS |
The ARM
01 is currently 510K Cleared
for Peracetic Acid/ Hydrogen Peroxide. |